- 0:00SPEAKER_01: Imagine authorizing your engineering team to push an update to a life support device
- 0:05SPEAKER_01: and the work order literally just says, clean up all code.
- 0:09SPEAKER_00: Oh, yeah.
- 0:10SPEAKER_00: Terrifying.
- 0:11SPEAKER_01: Right.
- 0:11SPEAKER_01: I mean, no record of what was actually changed.
- 0:14SPEAKER_01: No testing.
- 0:15SPEAKER_01: Just, you know, a blind software update pushed out into the wild.
- 0:19SPEAKER_00: Just crossing your fingers and hoping for the best.
- 0:21SPEAKER_01: Exactly.
- 0:21SPEAKER_01: And today we are doing a deep dive into an FDA warning letter
- 0:26SPEAKER_01: that shows exactly what happens when a medical device company treats safety like
- 0:32SPEAKER_01: a rough draft.
- 0:33SPEAKER_00: Which is unfortunately more common than you'd think.
- 0:35SPEAKER_01: Yeah.
- 0:35SPEAKER_01: And for you executives listening,
- 0:37SPEAKER_01: here is the absolute strategic bottom line
- 0:39SPEAKER_01: on quality management system risk governance.
- 0:43SPEAKER_01: Compliance is not a regulatory tax on your business.
- 0:46SPEAKER_00: Right.
- 0:46SPEAKER_00: It's not just a hoof to jump through.
- 0:47SPEAKER_01: It's routine operational systems design.
- 0:50SPEAKER_01: I mean, it as is what gives you speed to market,
- 0:53SPEAKER_01: predictable product performance,
- 0:54SPEAKER_01: and ultimately it protects your long term enterprise value.
- 0:58SPEAKER_00: Yeah, because when a leadership team views compliance
- 1:00SPEAKER_00: as this separate annoying hurdle rather than,
- 1:03SPEAKER_00: you know,
- 1:04SPEAKER_00: the foundation of their strategy, they literally lose control of their own product.
- 1:08SPEAKER_01: 100%.
- 1:09SPEAKER_00: I mean, the regulations we're going to unpack today aren't arbitrary rules invented
- 1:12SPEAKER_00: by bureaucrats.
- 1:14SPEAKER_00: They're codified engineering principles.
- 1:16SPEAKER_01: So I want to dig right into this because our source material today is an FDA warning letter.
- 1:22SPEAKER_01: It was issued on August 21st, 2026 to Lyon Street Medical doing business as Pensar Medical LLC.
- 1:30SPEAKER_00: Right.
- 1:30SPEAKER_00: They manufacture negative pressure wound therapy systems.
- 1:32SPEAKER_01: Yeah.
- 1:33SPEAKER_01: Devices like the Wound Pro,
- 1:34SPEAKER_01: the Wound Pro Plus A, and the MicroDoc, which is also referred to as the PocketDoc.
- 1:40SPEAKER_00: Yeah.
- 1:40SPEAKER_01: But before we get into the failures,
- 1:42SPEAKER_01: let's just establish what these devices actually do because you really need
- 1:45SPEAKER_01: to understand the stakes here.
- 1:47SPEAKER_00: Well, negative pressure wound therapy,
- 1:48SPEAKER_00: it's used for patients with really severe, often chronic wounds.
- 1:52SPEAKER_00: So think diabetic ulcers,
- 1:55SPEAKER_00: severe burns, or like large surgical incisions that just won't close properly.
- 2:00SPEAKER_01: Okay, so highly vulnerable patients.
- 2:02SPEAKER_00: Exactly.
- 2:02SPEAKER_00: And the system involves placing a specialized foam dressing directly into
- 2:05SPEAKER_00: the wound bed, sealing it with an airtight film and connecting it to a vacuum pump.
- 2:10SPEAKER_00: Right.
- 2:10SPEAKER_00: So the device applies continuous or intermittent negative pressure,
- 2:14SPEAKER_00: a vacuum effect basically,
- 2:15SPEAKER_00: to draw out fluid or exudate and infectious materials into a little canister.
- 2:20SPEAKER_01: So it physically pulls the edges of the wound together while removing the fluids
- 2:25SPEAKER_01: that cause infection.
- 2:27SPEAKER_01: And that promotes blood flow and tissue growth.
- 2:29SPEAKER_00: That is the goal.
- 2:30SPEAKER_01: Yeah.
- 2:30SPEAKER_00: And because it interacts directly with compromised tissue,
- 2:34SPEAKER_00: the mechanical precision of that pump,
- 2:36SPEAKER_00: the sterility of the packaging,
- 2:38SPEAKER_00: and the reliability of the software controlling the vacuum pressure,
- 2:41SPEAKER_00: it's all absolutely critical.
- 2:43SPEAKER_01: Because if it fails.
- 2:44SPEAKER_00: If the pressure drops, the wound just stops healing.
- 2:47SPEAKER_00: But if the pressure spikes, you can actually damage that fragile new tissue.
- 2:50SPEAKER_01: Which brings us to the mission for this deep dive.
- 2:53SPEAKER_01: We are going to explore how a failure to integrate risk management across
- 2:57SPEAKER_01: the entire product lifecycle led to a complete cascade of operational failures
- 3:02SPEAKER_01: for Pensar.
- 3:02SPEAKER_00: Right.
- 3:02SPEAKER_00: Looking at this through the lens of the QMSR,
- 3:05SPEAKER_00: the Quality Management System Regulation, and ISO 13485.
- 3:09SPEAKER_01: Yeah.
- 3:09SPEAKER_01: Specifically,
- 3:10SPEAKER_01: what happens when risk management isn't built into that initial manufacturing phase,
- 3:16SPEAKER_01: how software changes become incredibly dangerous without a baseline,
- 3:20SPEAKER_01: and why a broken KP system basically guarantees you'll repeat your mistakes.
- 3:25SPEAKER_00: Well, the best place to start is at the very beginning of a product's life cycle.
- 3:29SPEAKER_00: What the industry calls product realization.
- 3:31SPEAKER_01: Which is when it goes from a drawing to a real thing, right?
- 3:34SPEAKER_00: Exactly.
- 3:34SPEAKER_00: It's the phase where you transition a design on paper into
- 3:38SPEAKER_00: a physical product you can actually manufacture and sell.
- 3:42SPEAKER_00: And Clause 7.1 has a very clear rule here.
- 3:45SPEAKER_00: You must document risk management during this realization phase.
- 3:48SPEAKER_01: Right.
- 3:48SPEAKER_01: And according to the warning letter,
- 3:50SPEAKER_01: the FDA inspector looked at Pensar's design failure mode and effects analysis,
- 3:54SPEAKER_01: their DFMEA for the pocket dock device.
- 3:57SPEAKER_00: And what did they find?
- 3:58SPEAKER_01: Well, the document did identify distinct hazards,
- 4:01SPEAKER_01: but when it came to the actual risk control verifications,
- 4:04SPEAKER_01: you know, the proof that they mitigated those hazards.
- 4:06SPEAKER_00: Let me guess, they were missing.
- 4:08SPEAKER_01: Entries for software,
- 4:09SPEAKER_01: sterilization,
- 4:10SPEAKER_01: packaging, exudate management, they were all simply marked in progress.
- 4:14SPEAKER_00: Oh, wow.
- 4:15SPEAKER_00: In progress.
- 4:16SPEAKER_01: Yeah.
- 4:16SPEAKER_00: I mean, leaving a risk control marked in progress means you've identified a way
- 4:21SPEAKER_00: the device could potentially harm a patient,
- 4:23SPEAKER_00: but you just haven't bothered to prove your proposed solution actually works.
- 4:27SPEAKER_01: I am genuinely trying to wrap my head around this.
- 4:30SPEAKER_01: If I'm a CEO and I see in progress on my critical risk controls
- 4:34SPEAKER_01: for something like sterilization,
- 4:36SPEAKER_01: my first instinct is, okay, this product physically cannot leave the building.
- 4:41SPEAKER_00: Right.
- 4:41SPEAKER_00: It shouldn't.
- 4:42SPEAKER_01: Am I misunderstanding how FDA clearance works here?
- 4:45SPEAKER_01: Or did this company just totally bypass their own gatekeepers?
- 4:48SPEAKER_00: Well, it's actually a super common misconception about how the regulatory framework operates.
- 4:53SPEAKER_00: Oh, really?
- 4:54SPEAKER_00: Yeah.
- 4:54SPEAKER_00: The FDA clears the initial design of a device based on a pre-market submission,
- 4:59SPEAKER_00: assuming the data you provide is accurate.
- 5:01SPEAKER_00: But they don't, like, sit in your facility checking every piece of paper before a box ships.
- 5:05SPEAKER_01: Oh, so they're relying on you?
- 5:07SPEAKER_00: Completely.
- 5:08SPEAKER_00: They grant the clearance with the legal expectation
- 5:11SPEAKER_00: that your internal quality management system will act as the gatekeeper.
- 5:15SPEAKER_00: Pensar essentially dismantled their own gates.
- 5:17SPEAKER_01: You know what this reminds me of?
- 5:19SPEAKER_01: From an executive standpoint, it's almost like building a massive cargo ship.
- 5:24SPEAKER_00: Okay.
- 5:24SPEAKER_01: And you launch it into the harbor while
- 5:26SPEAKER_01: the hull's watertight doors are still just like a sketch on a notepad.
- 5:30SPEAKER_00: That is a perfect analogy.
- 5:32SPEAKER_01: You might float out of the harbor on a sunny day,
- 5:34SPEAKER_01: but the second you hit rough water, you are taking on water and going down.
- 5:37SPEAKER_01: I mean, how does an operation even function without a basic risk map?
- 5:43SPEAKER_00: Yeah.
- 5:43SPEAKER_00: Yeah.
- 6:05SPEAKER_00: But what is truly staggering is that for their Wound Pro product line,
- 6:09SPEAKER_00: the FDA found absolutely zero FMEAs of any type.
- 6:13SPEAKER_01: Wait, wait, zero.
- 6:15SPEAKER_01: No design FMEA, no process FMEA, no user FMEA.
- 6:18SPEAKER_00: Nothing was found to exist.
- 6:20SPEAKER_01: It's just nothing.
- 6:21SPEAKER_01: Wow.
- 6:22SPEAKER_00: And their own formal QMS process risk management procedure required all three
- 6:27SPEAKER_00: of those for every single product.
- 6:29SPEAKER_01: So they wrote a rule book for themselves and then entirely ignored it.
- 6:33SPEAKER_01: I really want to dig into the how of this.
- 6:36SPEAKER_01: Like, how does a company arrive at a point where
- 6:39SPEAKER_01: an entire product line has zero risk documentation?
- 6:43SPEAKER_00: Yeah.
- 6:43SPEAKER_01: Is it just sheer incompetence or is there some systemic operational flaw at play here?
- 6:48SPEAKER_00: Well, in many organizations that struggle with this, it really comes down to a siloed culture.
- 6:53SPEAKER_01: Okay, how so?
- 6:54SPEAKER_00: So research and development might design a product,
- 6:56SPEAKER_00: get the regulatory clearance,
- 6:58SPEAKER_00: and then basically toss the design over the wall to manufacturing.
- 7:01SPEAKER_01: Ah, I've seen that.
- 7:02SPEAKER_00: Right.
- 7:03SPEAKER_00: If R&D never finalized the risk verification and manufacturing is,
- 7:07SPEAKER_00: you know,
- 7:08SPEAKER_00: only incentivized by production quotas, the paperwork just sits in limbo forever.
- 7:11SPEAKER_01: And the warning letter also noted that when the FDA pointed this out,
- 7:15SPEAKER_01: Pensar's short-term correction was inadequate because they lacked
- 7:18SPEAKER_01: an interim risk mitigation plan.
- 7:20SPEAKER_00: Yes.
- 7:21SPEAKER_01: Let's translate that for the executives listening.
- 7:23SPEAKER_01: What exactly is an interim risk mitigation plan?
- 7:25SPEAKER_00: So if you're caught manufacturing a device without verified risk controls,
- 7:30SPEAKER_00: you can't just promise the FDA, hey,
- 7:32SPEAKER_00: we'll hire a consultant to write up the FMEA over the next six months
- 7:36SPEAKER_00: while you just keep shipping the product as usual.
- 7:38SPEAKER_01: Because you're still putting unverified devices on the market.
- 7:41SPEAKER_00: Exactly.
- 7:42SPEAKER_00: During those six months, you are still exposing patients to unverified risks.
- 7:47SPEAKER_00: An interim plan means putting immediate temporary guardrails in place.
- 7:51SPEAKER_01: Like what?
- 7:52SPEAKER_00: Like 100% manual inspection of every single unit.
- 7:55SPEAKER_00: Or even halting production entirely until the foundational paperwork is actually fixed.
- 8:00SPEAKER_00: And Pensar just didn't do this.
- 8:02SPEAKER_01: So if a company doesn't know what its baseline risks are,
- 8:05SPEAKER_01: and they don't have a verified map of how their product functions safely,
- 8:08SPEAKER_01: what happens when they inevitably need to change or update that product?
- 8:11SPEAKER_00: Well, that's where things are really dangerous.
- 8:13SPEAKER_01: Right.
- 8:14SPEAKER_01: That dynamic brings us to the next massive failure in this morning letter.
- 8:18SPEAKER_01: Which is Clause 7.3.9, the control of design and development changes.
- 8:23SPEAKER_01: Because without a baseline, engineering changes just become blind gambles.
- 8:27SPEAKER_00: Totally.
- 8:28SPEAKER_00: The standard requires any medical device manufacturer to adequately review,
- 8:33SPEAKER_00: verify, validate, and approve design changes before they're implemented.
- 8:38SPEAKER_00: You have to prove the change doesn't introduce a totally new hazard.
- 8:41SPEAKER_01: But because Pensar didn't have their baseline risk controls locked in,
- 8:45SPEAKER_01: their change management process was basically chaos.
- 8:48SPEAKER_00: Absolute chaos.
- 8:49SPEAKER_01: Let's look at the software side of this.
- 8:50SPEAKER_01: The FDGA found they implemented multiple software changes to the MicroDoc
- 8:55SPEAKER_01: and WoonPro Plus devices without any batch records confirming
- 8:59SPEAKER_01: the updates actually took place on the manufacturing line.
- 9:01SPEAKER_00: It's just wild.
- 9:02SPEAKER_01: For instance,
- 9:03SPEAKER_01: they had an engineering change order,
- 9:05SPEAKER_01: ECO 25039, for a WoonPro Plus firmware update, but zero proof it was installed.
- 9:11SPEAKER_00: And during the inspection,
- 9:12SPEAKER_00: the company's president actually confirmed
- 9:14SPEAKER_00: that no such batch record had been documented.
- 9:16SPEAKER_01: Unbelievable.
- 9:17SPEAKER_00: But the most glaring example of this chaos was ECO 25008.
- 9:21SPEAKER_00: The stated purpose of this engineering change order was to, and I quote, clean up all code.
- 9:27SPEAKER_01: Okay, I have to push back on this for a second.
- 9:29SPEAKER_01: Because in the broader tech world, right,
- 9:31SPEAKER_01: software engineers push massive code cleanups,
- 9:34SPEAKER_01: refactoring agile updates all the time.
- 9:37SPEAKER_01: It's standard practice to improve user experience.
- 9:40SPEAKER_00: Sure, in consumer tech.
- 9:41SPEAKER_01: Yeah.
- 9:42SPEAKER_01: So why is a general code cleanup such a cardinal sin in medical devices?
- 9:47SPEAKER_00: Well, the difference really lies in the consequences of a failure.
- 9:50SPEAKER_00: And just the interconnected nature of the software.
- 9:53SPEAKER_00: I mean, in consumer tech,
- 9:55SPEAKER_00: if a code cleanup introduces a bug,
- 9:56SPEAKER_00: an app crashes and you have to restart your phone.
- 9:59SPEAKER_01: Right.
- 9:59SPEAKER_01: Minor inconvenience.
- 10:00SPEAKER_00: But in a negative pressure wound therapy device,
- 10:03SPEAKER_00: if a code cleanup inadvertently changes, say, the timing of the vacuum sensor.
- 10:08SPEAKER_01: Oh, man.
- 10:09SPEAKER_00: Yeah.
- 10:09SPEAKER_00: The device might apply continuous high pressure to a fragile surgical wound.
- 10:14SPEAKER_00: That can cause severe hemorrhaging.
- 10:15SPEAKER_01: And the warning letter states that neither the CEO nor
- 10:18SPEAKER_01: the president could even explain what code changes were actually incorporated under
- 10:22SPEAKER_01: that cleanup all code order.
- 10:24SPEAKER_00: No verification record existed at all.
- 10:26SPEAKER_01: Hearing an engineering order to clean up all code with zero paper trail
- 10:31SPEAKER_01: that is terrifying for leadership, that is not a controlled change.
- 10:35SPEAKER_01: That is introducing a black box into your product line.
- 10:38SPEAKER_00: Exactly.
- 10:39SPEAKER_00: Which reveals a fundamental misunderstanding of regression testing
- 10:43SPEAKER_00: within their software validation procedure.
- 10:45SPEAKER_01: Walk us through regression testing.
- 10:47SPEAKER_00: So regression testing is the process of confirming that unchanged portions of
- 10:51SPEAKER_00: the software weren't inadvertently disrupted by a change you just made.
- 10:55SPEAKER_00: Software is deeply interdependent, right?
- 10:57SPEAKER_00: Right.
- 10:57SPEAKER_00: So a change in a module that controls,
- 11:00SPEAKER_00: let's say, the display screen could unexpectedly draw more power.
- 11:04SPEAKER_00: And that causes the module that controls the pump motor to starve and shut down.
- 11:08SPEAKER_01: Oh, wow.
- 11:09SPEAKER_00: If you aren't performing regression testing against a baseline risk map,
- 11:12SPEAKER_00: you are literally just crossing your fingers.
- 11:14SPEAKER_01: How does a team confidently drive product iteration and speed to market
- 11:18SPEAKER_01: if they cannot prove what code is in the device they just shipped?
- 11:21SPEAKER_00: They can't.
- 11:22SPEAKER_00: They also made a hardware or firmware change that they characterized as
- 11:27SPEAKER_00: a cosmetic noise change without documenting a risk assessment tied to
- 11:32SPEAKER_00: the original DFMEA.
- 11:34SPEAKER_01: They just said it was cosmetic.
- 11:35SPEAKER_00: They just declared it was unrelated to safety or effectiveness.
- 11:39SPEAKER_01: Let's break down the mechanics of why
- 11:40SPEAKER_01: a quote-unquote cosmetic noise change is never just cosmetic in a vacuum pump.
- 11:45SPEAKER_00: Okay, so to make a mechanical pump quieter,
- 11:47SPEAKER_00: you usually have to change the physical housing,
- 11:51SPEAKER_00: perhaps adding dampening materials,
- 11:53SPEAKER_00: or you change the voltage to lower the motor's RPMs.
- 11:56SPEAKER_01: And if you add dampening foam to the housing,
- 11:58SPEAKER_01: you might trap heat, which could cause the motor to overheat and fail.
- 12:02SPEAKER_00: Precisely.
- 12:03SPEAKER_00: Or if you lower the motor RPMs,
- 12:05SPEAKER_00: you might change the flow rate of the exudate being pulled from the wound.
- 12:09SPEAKER_00: You cannot simply declare a change safe without proving it against your risk map.
- 12:13SPEAKER_01: Every alteration has to be evaluated.
- 12:15SPEAKER_00: Every single one.
- 12:16SPEAKER_00: You have to ensure it doesn't introduce a new hazard.
- 12:19SPEAKER_00: Proper change governance isn't a roadblock designed to slow engineering down.
- 12:23SPEAKER_00: It's the very mechanism that allows you to innovate safely and quickly.
- 12:27SPEAKER_01: And this engineering chaos, it inevitably bleeds into regulatory trouble.
- 12:32SPEAKER_01: Specifically with corrections and removals,
- 12:36SPEAKER_01: What happened with the MicroDoc software update out in the field?
- 12:40SPEAKER_00: So Pensar realized they had a virgin issue with the MicroDoc software from multiple customers.
- 12:44SPEAKER_00: So they just conducted software reworks out in the field to get these customers on
- 12:48SPEAKER_00: the correct revision.
- 12:49SPEAKER_01: Okay.
- 12:50SPEAKER_00: Now, under 21 CFR Part 806,
- 12:53SPEAKER_00: fixing a software flaw in the field to reduce a risk to health is considered
- 12:56SPEAKER_00: a medical device correction.
- 12:58SPEAKER_00: And you are legally required to submit a report to the FDA within 10 working days.
- 13:03SPEAKER_01: Let me guess, they updated the software out in the wild without telling the agency.
- 13:06SPEAKER_00: They missed the 10-day window entirely.
- 13:08SPEAKER_00: As of the date of the warning letter, they still hadn't submitted it.
- 13:11SPEAKER_01: Man!
- 13:12SPEAKER_01: So we have unverified original designs combined
- 13:14SPEAKER_01: with unverified undocumented code changes and unreported field corrections.
- 13:18SPEAKER_01: It's a disaster.
- 13:19SPEAKER_01: The baseline is completely broken and the iteration process is blind.
- 13:24SPEAKER_01: In a healthy operational system,
- 13:25SPEAKER_01: when things break in the field,
- 13:28SPEAKER_01: internal feedback loops catch the issues before they spiral, right?
- 13:31SPEAKER_00: Usually, yes.
- 13:32SPEAKER_01: But this brings us to Clause 8.5.2 and MDRs.
- 13:37SPEAKER_01: Basically, corrective and preventive actions.
- 13:40SPEAKER_01: The final piece of this failure cascade.
- 13:42SPEAKER_01: Because at Pensar, the feedback loops were completely broken, cutting off the final safety net.
- 13:47SPEAKER_00: Your CAPIA system is supposed to be the engine of continuous improvement for your business.
- 13:52SPEAKER_00: When a failure occurs,
- 13:53SPEAKER_00: you investigate it,
- 13:55SPEAKER_00: you document the root cause,
- 13:56SPEAKER_00: implement a fix, and then you actually verify that the fix worked.
- 14:00SPEAKER_01: It's basic problem solving.
- 14:01SPEAKER_01: Yeah.
- 14:02SPEAKER_00: And the regulations explicitly require you to document these steps.
- 14:06SPEAKER_00: But the FDA reviewed five of Pensar's Cappier records and found this severe pattern
- 14:11SPEAKER_00: of blank fields.
- 14:12SPEAKER_01: I often think of CAPA as an enterprise's immune system.
- 14:15SPEAKER_01: You know,
- 14:15SPEAKER_01: when a pathogen,
- 14:16SPEAKER_01: like a defect or a field failure,
- 14:18SPEAKER_01: enters the body,
- 14:19SPEAKER_01: the white blood cells attack it,
- 14:20SPEAKER_01: learn from it, and create antibodies so it never harms the organization again.
- 14:23SPEAKER_00: I love that.
- 14:24SPEAKER_00: But Pinsir's immune system was completely offline.
- 14:27SPEAKER_00: CAPA 037, which was opened in January 2024, was entirely incomplete.
- 14:32SPEAKER_01: How incomplete?
- 14:33SPEAKER_00: The root cause section,
- 14:34SPEAKER_00: the corrective action sections,
- 14:35SPEAKER_00: and the effectiveness verification sections were all left completely blank.
- 14:39SPEAKER_01: How does that actually happen in a corporate environment?
- 14:42SPEAKER_01: Is it software fatigue like quality engineers just opening tickets in
- 14:47SPEAKER_01: a digital system to hit a geeky eye and then forgetting about them?
- 14:50SPEAKER_00: It's often a mix of template fatigue and honestly a culture of fear.
- 14:54SPEAKER_01: Oh, interesting.
- 14:55SPEAKER_00: Yeah.
- 14:56SPEAKER_00: When compliance is viewed as just a paperwork exercise,
- 14:59SPEAKER_00: an engineer might open a KPI just so they can tell management,
- 15:02SPEAKER_00: hey, I opened a ticket.
- 15:03SPEAKER_01: Right.
- 15:04SPEAKER_00: Box check.
- 15:05SPEAKER_00: Exactly.
- 15:06SPEAKER_00: Another issue is that documenting a root cause feels like admitting liability to some people.
- 15:12SPEAKER_00: If an organization doesn't have a culture of psychological safety,
- 15:15SPEAKER_00: employees will literally leave fields blank rather than put in writing that
- 15:20SPEAKER_00: the engineering team messed up.
- 15:21SPEAKER_01: Wow.
- 15:22SPEAKER_00: We saw this with KPAR Series 9,
- 15:24SPEAKER_00: which contained only unfilled placeholder language that carried over from
- 15:28SPEAKER_00: a template.
- 15:28SPEAKER_01: So they just opened a digital template, saved it, and never did the work.
- 15:31SPEAKER_01: Seeing a completely blank KPAR is like your white blood cells clocking out for lunch
- 15:36SPEAKER_01: during a severe infection.
- 15:38SPEAKER_01: No, and that is where the entire QMS fractures.
- 15:40SPEAKER_01: KPA 036 noted that a corrective action was taken.
- 15:41SPEAKER_01: But that action was never incorporated back into the firm's risk analysis,
- 15:44SPEAKER_01: and the effectiveness fields were left blank.
- 16:01SPEAKER_01: This is a critical point.
- 16:02SPEAKER_01: If we want to build resilient organizations,
- 16:05SPEAKER_01: how do you practically integrate risk analysis back into a KPA system
- 16:08SPEAKER_01: so it actually works?
- 16:10SPEAKER_00: Well, it has to be a continuous closed loop system.
- 16:13SPEAKER_00: When a failure occurs, your investigation must ask one vital question.
- 16:17SPEAKER_00: Did we predict this failure in our original FMEA?
- 16:20SPEAKER_01: Okay.
- 16:21SPEAKER_00: If the answer is yes, then you have to ask why your predicted risk control failed.
- 16:25SPEAKER_00: But if the answer is no,
- 16:26SPEAKER_00: then your FMEA is incomplete and you have to update it
- 16:29SPEAKER_00: with this newly discovered hazard.
- 16:30SPEAKER_01: But you can only trigger a KPAY if your internal definitions for recognizing
- 16:35SPEAKER_01: a problem are actually clear in the first place.
- 16:38SPEAKER_01: Which brings us to their medical device reporting, or MDR, failures.
- 16:42SPEAKER_01: Right.
- 16:43SPEAKER_01: Because an MDR is often the trigger for a major KPAY.
- 16:47SPEAKER_00: The regulation dictates that you have to report an event to the FDA
- 16:50SPEAKER_00: if information reasonably suggests your device caused or contributed to a death
- 16:55SPEAKER_00: or serious injury.
- 16:57SPEAKER_00: But Pensar's internal procedure for MDRs didn't even include an adequate definition
- 17:03SPEAKER_00: of what reasonably suggests means.
- 17:05SPEAKER_01: Let's look at the psychology of that, because reasonably suggests wasn't clearly defined.
- 17:09SPEAKER_01: Imagine a quality engineer sitting in a cubicle gets a complaint about a wound infection.
- 17:14SPEAKER_01: Without a strict definition, human nature takes over, right?
- 17:17SPEAKER_01: They downplay it.
- 17:18SPEAKER_01: They think infections happen all the time.
- 17:19SPEAKER_01: The patient was already sick.
- 17:21SPEAKER_01: It's just a device malfunction.
- 17:22SPEAKER_00: Yeah, that happens a lot.
- 17:23SPEAKER_01: They don't realize the regulatory standard requires them to assume
- 17:26SPEAKER_01: the device contributed to the injury until they can definitively prove otherwise.
- 17:31SPEAKER_00: And that lack of definition led directly to the mishandling of complaint C-029.
- 17:37SPEAKER_00: A patient using Pensar's negative pressure system sustained an infection at
- 17:41SPEAKER_00: the wound site, specifically purulent drainage.
- 17:44SPEAKER_01: Which means it was actively weeping infectious fluid.
- 17:47SPEAKER_00: Exactly.
- 17:48SPEAKER_00: The situation escalated to the point where the patient required medical
- 17:51SPEAKER_00: and surgical intervention to remove retained foam from the wound
- 17:55SPEAKER_00: to preclude permanent impairment.
- 17:56SPEAKER_01: Just so you listening understand the medical mechanics here,
- 18:00SPEAKER_01: the foam dressing from the device was left inside the patient's wound.
- 18:03SPEAKER_01: And when that happens,
- 18:04SPEAKER_01: the body's natural granulation tissue actually starts growing into
- 18:08SPEAKER_01: the porous structure of the foam.
- 18:10SPEAKER_00: It's awful.
- 18:10SPEAKER_01: You can't just pull it out like a Band-Aid.
- 18:12SPEAKER_01: A surgeon has to go in with a scalpel and physically excise the foam and the new tissue.
- 18:16SPEAKER_00: It is a severe event.
- 18:18SPEAKER_00: That is the very definition of a serious injury under the regulations.
- 18:22SPEAKER_00: Now, Pensar became aware of this event on May 1st, 2024.
- 18:26SPEAKER_00: They failed to submit the required MDR within the 30 calendar day window.
- 18:30SPEAKER_01: Typical.
- 18:31SPEAKER_00: And when they finally did submit it,
- 18:32SPEAKER_00: they incorrectly categorized it as a malfunction report instead of
- 18:35SPEAKER_00: a serious injury report.
- 18:37SPEAKER_01: How does a quality engineer downgrade a surgical excision to a mere malfunction?
- 18:42SPEAKER_00: Well, because the procedure didn't define reasonably suggests,
- 18:46SPEAKER_00: they likely convinced themselves that the infection caused the surgery,
- 18:50SPEAKER_00: not the retained foam,
- 18:52SPEAKER_00: completely missing the fact that the device was inextricably linked to the event.
- 18:56SPEAKER_01: Which completely misrepresents the severity of the issue to the regulatory agencies
- 19:01SPEAKER_01: and, frankly, to their own internal management teams looking at the data.
- 19:04SPEAKER_00: Right.
- 19:05SPEAKER_01: If the executive team sees a dashboard with malfunctions, they don't allocate resources.
- 19:10SPEAKER_01: Right.
- 19:11SPEAKER_01: If they see serious injuries requiring surgery, they stop the assembly line.
- 19:15SPEAKER_00: When you connect all these dots, you really see the breathtaking cascade of systemic failure.
- 19:20SPEAKER_00: A serious injury occurs.
- 19:22SPEAKER_00: It is miscategorized and delayed in reporting because the procedures lack basic definitions.
- 19:27SPEAKER_00: Because the reporting is flawed, the severity isn't recognized by leadership.
- 19:31SPEAKER_00: Therefore,
- 19:32SPEAKER_00: the resulting KPAs are opened but left as blank templates or closed
- 19:37SPEAKER_00: without effectiveness checks.
- 19:39SPEAKER_00: It all stems from viewing risk management as an isolated theoretical exercise rather
- 19:42SPEAKER_00: than the living, breathing operational system it is supposed to be.
- 19:44SPEAKER_00: 100%.
- 19:59SPEAKER_01: Leaving KPAs blank or disconnected from risk analysis basically guarantees
- 20:03SPEAKER_01: that you will repeat your failures.
- 20:05SPEAKER_01: You will keep shipping devices with unverified software changes,
- 20:09SPEAKER_01: and you will keep having retained foam incidents because you never closed the loop.
- 20:12SPEAKER_00: None of these FDA citations are about government overreach or nitpicking typos.
- 20:17SPEAKER_00: They're about a company that lost total control of its engineering and quality feedback loops.
- 20:23SPEAKER_00: A solid, interconnected quality management system is a strategic asset.
- 20:28SPEAKER_00: When your risk controls are verified before launch,
- 20:30SPEAKER_00: when your software changes are regression tested against the baseline,
- 20:34SPEAKER_00: and when your corrective actions actually update your corporate knowledge.
- 20:37SPEAKER_01: Things just work.
- 20:38SPEAKER_00: Yes.
- 20:39SPEAKER_00: Your business runs with incredible efficiency.
- 20:42SPEAKER_00: You can innovate faster because you know exactly where the guardrails are.
- 20:46SPEAKER_01: That is the mindset shift that separates industry leaders from companies drowning
- 20:50SPEAKER_01: in compliance issues.
- 20:52SPEAKER_01: It is about maintaining absolute,
- 20:54SPEAKER_01: verifiable control over your product's lifecycle to drive enterprise value.
- 21:00SPEAKER_00: And Pensar serves as a stark reminder of what happens when
- 21:04SPEAKER_00: that operational control is entirely abandoned.
- 21:08SPEAKER_01: Oh, wow.
- 21:08SPEAKER_01: Yeah.
- 21:33SPEAKER_01: If
- 21:33SPEAKER_01: an organization's quality management system cannot even govern static human written
- 21:38SPEAKER_01: code today, how in the world will it handle the autonomous engineering of tomorrow?