- 0:00Speaker 1: Welcome to the deep dive. Today. We're undertaking a really
- 0:03Speaker 1: critical mission. We're going to distill the standards of care
- 0:07Speaker 1: in laboratory diagnosis.
- 0:09Speaker 2: That's right.
- 0:09Speaker 1: We've been digging into some key clinical chemical sources to
- 0:12Speaker 1: give you a genuine shortcut to understanding where this high
- 0:16Speaker 1: stake system of medical testing most often fails.
- 0:19Speaker 2: And this is just absolutely foundational material for anyone in healthcare.
- 0:23Speaker 2: Our sources all guide us with that primary ethical principle
- 0:27Speaker 2: you know, above all, do no harm, right, And what's
- 0:30Speaker 2: so crucial to understand is that minimizing diagnostic error it
- 0:34Speaker 2: means looking at the entire process. It's not just about
- 0:37Speaker 2: the machines.
- 0:38Speaker 1: That's where the real complexity is, isn't it. We all
- 0:40Speaker 1: have this picture of sleek automated instruments, but the sources
- 0:44Speaker 1: make it clear that the automation is well, it's irrelevant
- 0:47Speaker 1: if the initial sample is compromised. So what's the true
- 0:50Speaker 1: focus of error prevention here?
- 0:52Speaker 2: It's the pre analytical and the post analytical phases. Most
- 0:55Speaker 2: of these failures actually happen outside the laboratory.
- 0:58Speaker 1: Walls, so when the blood is being collected or when
- 1:01Speaker 1: the results are reported, exactly when.
- 1:03Speaker 2: It's being collected, transported, or interpreted. True quality management requires
- 1:09Speaker 2: aggressive interdisciplinary collaboration. We're talking between nursing, phlebotomy, informatics, and
- 1:16Speaker 2: the lab staff.
- 1:17Speaker 1: So we're going to cover some of those common high
- 1:19Speaker 1: frequency mistakes.
- 1:21Speaker 2: The ones made daily across thousands of institutions.
- 1:24Speaker 1: Yes, okay, let's start with segment one, the critical foundation.
- 1:27Speaker 1: This is all about specimen integrity and pre analytical failure.
- 1:31Speaker 2: The very beginning of the process and the.
- 1:33Speaker 1: First layer of defense is purely human. It's just verifying identity.
- 1:37Speaker 2: And the standard of care here is well, it's non negotiable.
- 1:41Speaker 2: It's designed to completely eliminate mixups. You must use a
- 1:45Speaker 2: minimum of two unique patient identifiers.
- 1:48Speaker 1: So full name, date of birth, something like that.
- 1:50Speaker 2: Usually full name, day of birth, or the medical record
- 1:52Speaker 2: number the MRN. And here's the kicker. Labeling of the
- 1:56Speaker 2: collection twos has to happen immediately after collection, in the
- 1:58Speaker 2: patient's presence.
- 2:00Speaker 1: So why is that last rule so strict? I mean,
- 2:02Speaker 1: why can't a phlebotomist just label the tubes once they
- 2:05Speaker 1: get back to the workstation.
- 2:07Speaker 2: Because pre labeling is a prime opportunity for catastrophic error.
- 2:11Speaker 1: It just is a mix up waiting to happen.
- 2:14Speaker 2: Absolutely. Imagine a philebotomists prepares a whole rack of tubes
- 2:18Speaker 2: for three different patients that you distracted. They come back
- 2:22Speaker 2: and they use the wrong two on a patient.
- 2:24Speaker 1: And that result will now follow that patient forever indefinitely.
- 2:28Speaker 2: The requirement to label immediately while you're confirming the identifiers
- 2:32Speaker 2: with the patient is really the only robust safeguard we
- 2:35Speaker 2: have against misidentification.
- 2:37Speaker 1: The sources have some great anecdotes that illustrate how policies
- 2:40Speaker 1: that seem like bureaucratic hurdles are actually critical safety nets. Yeah,
- 2:44Speaker 1: that delayed blood gas case is a perfect example.
- 2:47Speaker 2: It is so a nurse collected a blood gas sample
- 2:51Speaker 2: but failed to label it right there. So the laboratory
- 2:54Speaker 2: correctly rejected the specimen based on policy. So a second
- 2:57Speaker 2: collection was attempted, but that sample it arrived at the
- 3:00Speaker 2: lab delayed. We're talking more than forty five minutes after collection,
- 3:04Speaker 2: and it was at room temperature, not on ice.
- 3:06Speaker 1: And what happened to the results then, Well, for highly.
- 3:09Speaker 2: Volable analytes like blood gases, pH ionized calcium, the sample
- 3:15Speaker 2: just degrades incredibly rapidly with time and.
- 3:18Speaker 1: Temperature, so the data is garbage.
- 3:19Speaker 2: It's completely unreliable. The lab had to cancel those results,
- 3:23Speaker 2: which then forced a healthcare team to collect a fourth sample,
- 3:26Speaker 2: a fourth one, a fourth, And while the ordering physician
- 3:29Speaker 2: was you know, irate about the obstructive policy, the lab's
- 3:32Speaker 2: strict adherence prevented the release of misleading data, data that
- 3:36Speaker 2: could have driven inappropriate, even fatal medical intervention.
- 3:40Speaker 1: So those policies, they really are there to protect the
- 3:42Speaker 1: patient from bad data.
- 3:44Speaker 2: That's the only purpose.
- 3:45Speaker 1: Okay, Moving beyond identification, collection errors can ruin the chemical
- 3:49Speaker 1: integrity of the specimen itself. Let's talk about additive contamination.
- 3:53Speaker 2: The simplest case is just using the wrong tube. Imagine
- 3:56Speaker 2: a test requires a red top tube that means no additive,
- 4:00Speaker 2: but instead the lab gets a purple.
- 4:01Speaker 1: Top tube which has edta in it right.
- 4:03Speaker 2: Which contains dta. And Edta is a powerful anticoagulant that
- 4:08Speaker 2: works by chelating or binding calcium ions.
- 4:11Speaker 1: So any test that needs calcium is going to be
- 4:13Speaker 1: thrown off.
- 4:14Speaker 2: Any assay that relies on free calcium coagulation studies, certain
- 4:19Speaker 2: trace metal tests, even a free T four immuno assey,
- 4:24Speaker 2: the edta interference just renders the result chemically invalid. It's
- 4:29Speaker 2: an immediate recollection, no other option.
- 4:31Speaker 1: That's pure chemistry preventing the analysis. Now what about that
- 4:35Speaker 1: pre analytical failure of using an indwelling IV catheter for
- 4:39Speaker 1: collection that's known to cause dilution contamination.
- 4:43Speaker 2: Yeah, you hit on a really high risk area. Since
- 4:45Speaker 2: collecting blood through an IV line can lead to unreliable results,
- 4:48Speaker 2: the process has to be strictly controlled.
- 4:50Speaker 1: So what's the standard of care to mitigate that risk?
- 4:53Speaker 2: Okay, So if a clinician determines that collection through an
- 4:56Speaker 2: indwelling line is absolutely necessary, maybe because of truly poor
- 5:00Speaker 2: vascular access or a high bleeding risk, three specific requirements
- 5:04Speaker 2: must be met.
- 5:04Speaker 1: Okay, what are they?
- 5:05Speaker 2: First, it must be based on a physician order, one
- 5:07Speaker 2: that specifically acknowledges the risks. Second, it can only be
- 5:11Speaker 2: performed on patients where poor access justifies that risk. That
- 5:14Speaker 2: the third and third, the result must include a specimen comment,
- 5:17Speaker 2: something that notes it is collected through line.
- 5:20Speaker 1: Why is that so important?
- 5:21Speaker 2: Because it alerts the ordering provider to the potential for
- 5:24Speaker 2: dilution or contamination. It helps them troubleshit the result, It
- 5:29Speaker 2: gives context.
- 5:29Speaker 1: So those three steps ensure that if contamination does happen,
- 5:33Speaker 1: the result isn't just treated as gospel percise. Speaking of
- 5:37Speaker 1: compromised samples, what about temperature abuse using outdoor specimen drop
- 5:42Speaker 1: boxes seems like a minor logistics issue, but it causes huge.
- 5:45Speaker 2: Problems, a massive problem, especially for outreach programs. If a
- 5:48Speaker 2: sample freezes, it causes hemolysis.
- 5:51Speaker 1: The red blood cells burst.
- 5:52Speaker 2: They burst, and that releases massive amounts of intracellular components
- 5:56Speaker 2: into the serum. So it artificially elevates potassium, ldlack date, dehydrogenase,
- 6:02Speaker 2: and certain enzymes.
- 6:03Speaker 1: And what about heat?
- 6:04Speaker 2: If the sample overheats, accelerated cellular metabolism changes glucose and
- 6:08Speaker 2: pH levels. The lab has to provide clear, environmentally controlled
- 6:13Speaker 2: recommendations for transport. It's the only way to ensure analyte stability.
- 6:17Speaker 1: Okay, let's transition into segment two. Inside the lab, we're
- 6:20Speaker 1: focusing on analytical pitfalls, shortcuts, and quality control. Now, the
- 6:25Speaker 1: quality control process is the lab's internal self check, and
- 6:29Speaker 1: the source describes a common and pretty dangerous shortcut that
- 6:33Speaker 1: technologists sometimes take.
- 6:35Speaker 2: This is the QC failure trap. A core requirement is
- 6:38Speaker 2: running QC samples at two different concentrations to verify that
- 6:42Speaker 2: the instrument is accurate across its whole measurement range.
- 6:45Speaker 1: Okay, so what's the shortcut.
- 6:47Speaker 2: The shortcut happens when it control fails and the technologist
- 6:50Speaker 2: just repeatedly runs that same sample maybe five times until
- 6:54Speaker 2: it randomly, by chance, falls just inside the acceptable.
- 6:58Speaker 1: Range, and then they release patient results.
- 7:00Speaker 2: Then they release patient results. It's essentially trying to hide
- 7:02Speaker 2: a system flaw.
- 7:03Speaker 1: So what should they be doing instead?
- 7:05Speaker 2: They need to pause. They need to stop. If QC
- 7:09Speaker 2: samples at two concentrations fail, especially if they both fail
- 7:12Speaker 2: on the same side of the mean, that signals a
- 7:15Speaker 2: systemic failure.
- 7:16Speaker 1: It's not random, It is.
- 7:17Speaker 2: Not random chance. It demands immediate instrument troubleshooting, you know,
- 7:22Speaker 2: checking for reagent degradation, recalibrating, maybe checking the light source.
- 7:26Speaker 2: Repetition is never a substitute for fixing the root cause.
- 7:29Speaker 1: Now. Point of care testing POCT is often marketed as
- 7:32Speaker 1: a solution to increase efficiency, but the sources show it
- 7:36Speaker 1: can actually create more management complexity.
- 7:38Speaker 2: It does because POCT implementation often tries to solve complex
- 7:42Speaker 2: scheduling or staffing failures rather than being driven by pure
- 7:46Speaker 2: medical necessity.
- 7:47Speaker 1: I think I know the case you're referring to.
- 7:49Speaker 2: The cardiovascular unit case, they implemented POCT for electrolytes and
- 7:54Speaker 2: it increased delays.
- 7:56Speaker 1: Why how did that happen?
- 7:57Speaker 2: Because the staff just added the POCT testing burden, all
- 8:01Speaker 2: the complex QC, the documentation proficiency testing onto an already
- 8:05Speaker 2: uncoordinated system. The standard is clear, POCT must be driven
- 8:10Speaker 2: by an immediate patient medical need, not as a shortcut
- 8:14Speaker 2: to bypass established system management failures.
- 8:16Speaker 1: And there are serious regulatory requirements for the operators of
- 8:19Speaker 1: these decentralized devices too. I mean regarding competency.
- 8:22Speaker 2: Yes, competency assessment has to be rigorous and specific to
- 8:26Speaker 2: the device. You can't just assume a general laboratory certification
- 8:29Speaker 2: covers it.
- 8:29Speaker 1: The sources highlight it a really shocking scenario with this the.
- 8:33Speaker 2: Legally blind operator attempting to run a POCT device. Yes,
- 8:38Speaker 2: this operator because they had difficulty reading the display and
- 8:41Speaker 2: the keypad was making serious identification errors. A supervisor has
- 8:46Speaker 2: to verify the operator's ability to safely and correctly operate
- 8:50Speaker 2: the specific device they're assigned to, and if they can't,
- 8:52Speaker 2: their privileges must be suspended immediately, no question.
- 8:55Speaker 1: Before we move on. There's a critical safety issue. Tied
- 8:58Speaker 1: to the analytical process. We need to cover first the
- 9:00Speaker 1: mechanical hazard centrifagation.
- 9:03Speaker 2: Centrification is it's deceptively hazardous. We saw the case where
- 9:08Speaker 2: an incorrect RPM calculation caused a high speed centrifuge to
- 9:13Speaker 2: run too fast, generate in a lot of friction, a
- 9:15Speaker 2: ton of fiction. This caused the temperature inside the rotor
- 9:18Speaker 2: chamber to rise dramatically up to forty five fifty degree celsius,
- 9:22Speaker 2: and that compromised all the heat sensitive analytes in the samples.
- 9:25Speaker 1: So proper calculation is essential.
- 9:27Speaker 2: Calculation of g force and careful monitoring of temperature it's
- 9:31Speaker 2: essential to prevent that kind of degradation and sample loss.
- 9:34Speaker 1: And then there's the severe laboratory safety hazard involving improperly
- 9:39Speaker 1: handled biological waste. The explosive specimen. Oh yeah, you have
- 9:43Speaker 1: to tell the listener about the fecal fast specimen.
- 9:46Speaker 2: This case is famous or maybe infamous within lab safety circles.
- 9:51Speaker 2: A specimen for fecal fat analysis was collected in an
- 9:54Speaker 2: unvented paint can.
- 9:55Speaker 1: An unvented paint can.
- 9:57Speaker 2: Exactly, and because it was left in a hot outdoor
- 9:59Speaker 2: drop over a long holiday weekend, bacterial gas production just
- 10:03Speaker 2: went into overdrive, generating immense pressure inside the can.
- 10:06Speaker 1: I can see where this is going.
- 10:08Speaker 2: When the can was brought into the processing area, it
- 10:10Speaker 2: violently exploded. You're king, No, the biological material was aerosolized.
- 10:15Speaker 2: It was embedded into every single surface of the processing area.
- 10:18Speaker 2: They had a closed room down and sand blasted.
- 10:20Speaker 1: Wait, sand blasted, sand blasted.
- 10:22Speaker 2: To ensure complete biohazard remediation. The standard of care is
- 10:26Speaker 2: absolutely clear. High gas producing specimens, especially those transported over
- 10:31Speaker 2: extended periods, must use vented containers period.
- 10:36Speaker 1: Wow. That illustrates the stakes perfectly. Okay, let's move into
- 10:40Speaker 1: segment three. Informatics, we're covering the LIS, the EMR, and
- 10:44Speaker 1: the post analytical breakdown. So for our listeners who might
- 10:48Speaker 1: not be familiar with that jargon, what are we talking
- 10:50Speaker 1: about here?
- 10:51Speaker 2: So the LIS is the Laboratory Information system, that's the
- 10:54Speaker 2: specialized software where testing is ordered and results in processed.
- 10:57Speaker 1: And the EMR.
- 10:58Speaker 2: The EMR is the electroconic medical record. It's the patient's
- 11:01Speaker 2: permanent chart where clinicians view those results. And failures at
- 11:06Speaker 2: this interface are incredibly common and.
- 11:08Speaker 1: One of the simplest failures starts right at the beginning
- 11:11Speaker 1: test naming confusion which leads to overordering.
- 11:15Speaker 2: This happens all the time when new test methodologies are introduced.
- 11:18Speaker 2: For example, if you introduce a high sensitivity.
- 11:20Speaker 1: CRP which is for cardiac risk.
- 11:22Speaker 2: Correct the naming in the LIS has to carefully distinguish
- 11:25Speaker 2: it from the standard CRP used for inflammation or sepsis.
- 11:29Speaker 2: If the names are confusing, clinicians will misapply the test, and.
- 11:32Speaker 1: That overordering leads directly into what the source is call
- 11:35Speaker 1: the incidental finding cascade. Tell us about the resident who
- 11:39Speaker 1: ordered the entire thyroid panel.
- 11:41Speaker 2: Right, So this resident, unsure of the correct way to
- 11:44Speaker 2: evaluate hypothyroidism, order the whole suite TSH FT four the
- 11:49Speaker 2: rarely used T three.
- 11:50Speaker 1: Are you just everything when the standard is sequential?
- 11:53Speaker 2: The standard is the sequential strategy. You start with TSH
- 11:56Speaker 2: and then you reflex to FT four only if the
- 11:58Speaker 2: TSH is abnormal, clinically sound, and cost effective. By ordering
- 12:02Speaker 2: the whole panel, they found a detectable level of thyroid
- 12:05Speaker 2: binding inhibitory immune globulin.
- 12:07Speaker 1: So finding this immunoglobulin, even if it's not clinically significant,
- 12:11Speaker 1: it forces further action.
- 12:13Speaker 2: Right precisely, even though the tier wasn't high enough to
- 12:16Speaker 2: meet diagnostic criteria. Just the fact that an abnormal result appeared.
- 12:20Speaker 2: It caused an additional unnecessary work up for autoimmune disease.
- 12:23Speaker 1: So overutilization increases costs and can be misleading.
- 12:27Speaker 2: It absolutely burdens the patient and the system.
- 12:29Speaker 1: The final post analytical failures happen when results are viewed,
- 12:32Speaker 1: and the case of the suppressed specimen comment in the
- 12:34Speaker 1: EMR is just terrifying.
- 12:36Speaker 2: This is a huge design flaw in how EMRs can
- 12:38Speaker 2: present data. A critical care resident was viewing a low
- 12:41Speaker 2: potassium result two point nine milli mall in a tabular
- 12:44Speaker 2: summ review.
- 12:45Speaker 1: It's a table of numbers exactly.
- 12:47Speaker 2: And crucially, the specimen comment which stated specimen potentially contaminated
- 12:52Speaker 2: drawn above PYV line was hidden. It was suppressed, or
- 12:57Speaker 2: it required a separate click to even view it.
- 13:00Speaker 1: The resident was treating that low potassium as a genuine
- 13:02Speaker 1: finding that required intervention, exactly.
- 13:05Speaker 2: The standard of care demands that EMR systems actively draw
- 13:08Speaker 2: the user's attention to any critical comment associated with the
- 13:12Speaker 2: lab result, especially if the comment relates to specimen integrity.
- 13:16Speaker 2: Hiding those vital details risks catastrophic misinterpretation and Sometimes.
- 13:21Speaker 1: The interface failure isn't suppression, it's outright stage corruption.
- 13:24Speaker 2: Interface failures require ongoing monitoring, not just a one time verification.
- 13:28Speaker 2: We've seen data corruption where a CAA one twenty five
- 13:30Speaker 2: result over one hundred thousand was just displayed as too
- 13:33Speaker 2: big because the legacy interface couldn't handle six digits.
- 13:36Speaker 1: That's bad.
- 13:37Speaker 2: Even worse our decimal errors. A truly high stakes potassium
- 13:40Speaker 2: value of four point zero milli mol was displayed as
- 13:43Speaker 2: point four. Imagine the clinical action taken based on that
- 13:46Speaker 2: decimal shift. It's an immediate threat to life.
- 13:49Speaker 1: Okay, let's move to segment four specialized context, starting with
- 13:53Speaker 1: therapeutic drug monitoring and toxicology. These areas seem to involve
- 13:57Speaker 1: the highest level of patient deception.
- 13:59Speaker 2: They really do so. TDM involves monitoring drug levels things
- 14:03Speaker 2: like vancomycin or lithium, while toxicology screens for drugs of
- 14:07Speaker 2: abuse and patients use three main methods to generate false negatives,
- 14:12Speaker 2: which are dilution so drinking excessive water or using diuretics,
- 14:16Speaker 2: substitution using synthetic or borrowed urine, and adulteration.
- 14:20Speaker 1: How sophisticated is this adulteration?
- 14:23Speaker 2: Very they use everything from dish soap to acids. Adulterance
- 14:27Speaker 2: work by interfering with the ammino assay. For example, some
- 14:30Speaker 2: common household acids will cause a structural change in the
- 14:32Speaker 2: antibodies used in the assey.
- 14:34Speaker 1: Leading to a false negative.
- 14:35Speaker 2: Right. But conversely, some adulterants, like certain oxidants, can actually
- 14:39Speaker 2: cause a false positive by mimicking the analyte. That's why
- 14:42Speaker 2: the lab has to check integrity indicators like creatinine, osmolality.
- 14:46Speaker 1: And pH and the physician has to be aware that
- 14:48Speaker 1: even legal over the counter drugs can create false positives.
- 14:52Speaker 2: Absolutely. The clinical standard is that you should rarely, if ever,
- 14:57Speaker 2: treat a patient solely based on a screening amino assay result.
- 15:00Speaker 2: Common cold medicines with epandrine can cross react with amphetamine assays.
- 15:04Speaker 1: Okay, any other examples.
- 15:06Speaker 2: Yeah, dextromethorphan, which is in a lot of cough syrups,
- 15:09Speaker 2: can flag a false positive for PCP assays. Confirmatory testing,
- 15:13Speaker 2: usually with a highly specific method like GCMs, is essential
- 15:17Speaker 2: to rule these out.
- 15:18Speaker 1: Finally, let's revisit the critical value responsibility. This is governed
- 15:22Speaker 1: by specific regulations like CLIA. What's the non negotiable legal
- 15:27Speaker 1: imperative Here.
- 15:28Speaker 2: Critical values those results that represent an immediate life threatening situation.
- 15:33Speaker 2: They demand immediate interruptive contact. The lab must reach.
- 15:37Speaker 1: A live person, usually the ordering physician or their.
- 15:39Speaker 2: Dizzy knee, and crucially, that person must then verbally read
- 15:43Speaker 2: back the result to the technologists to confirm accurate transmission
- 15:46Speaker 2: and comprehension. It's a legal chain of custody for information.
- 15:49Speaker 1: We saw the friction that can happen when physicians try
- 15:52Speaker 1: to skirt this, like in the Vanko mice in.
- 15:53Speaker 2: Case that case, Yeah, it involved the Vanke micein test
- 15:57Speaker 2: ordered by a physician as a courtesy for a patient
- 16:00Speaker 2: who was already discharged. When the lab called with a
- 16:03Speaker 2: critical high value, the physician refused.
- 16:06Speaker 1: The call, claiming it wasn't his responsibility.
- 16:09Speaker 2: Right, it's a clear breach of professional expectation. Ordering physicians
- 16:13Speaker 2: must legally and professionally take responsibility for a follow up
- 16:17Speaker 2: on tests ordered under their name.
- 16:19Speaker 1: And what about outreach clients asking to leave messages.
- 16:23Speaker 2: A lab cannot legally comply with a request to leave
- 16:25Speaker 2: critical results on an answer machine. They have to maintain
- 16:28Speaker 2: that immediate live communication link.
- 16:31Speaker 1: It's required, okay, So to synthesize everything we've discussed, I'd.
- 16:35Speaker 2: Say the quality management and clinical chemistry. It extends so
- 16:38Speaker 2: far beyond just the analytical instrument. It requires continuous vigilance
- 16:42Speaker 2: across the entire system from start to finish, from the
- 16:45Speaker 2: phlebotomist ensuring two unique identifiers are used, to the technologists
- 16:49Speaker 2: troubleshooting a systemic QC failure, all the way to the
- 16:52Speaker 2: EMR that's actively displaying necessary specimen comments. True effectiveness depends
- 16:58Speaker 2: on collaboration, communication and constantly questioning assumptions.
- 17:02Speaker 1: That provides a really comprehensive picture. So to reinforce the
- 17:06Speaker 1: standards of care for you, our learner, let's end with
- 17:09Speaker 1: a short review exercise based on one of the pitfalls
- 17:12Speaker 1: we covered earlier. Excellent idea, we discuss the high risk
- 17:15Speaker 1: of contamination when collecting blood through an indwelling ABEA cafter So,
- 17:19Speaker 1: based on our discussion, what three specific requirements must be met.
- 17:23Speaker 1: If a clinician determines that collecting a specimen through an
- 17:26Speaker 1: induling catheter is absolutely necessary.
- 17:30Speaker 2: Think about the steps required to validate that risk. The
- 17:33Speaker 2: three requirements we talked about are, First, the collection has
- 17:37Speaker 2: to be based on a specific physician order that acknowledges
- 17:39Speaker 2: the risk. Second, it can only be done on patients
- 17:43Speaker 2: with truly compromised venus access or high bleeding risk. And third,
- 17:48Speaker 2: the laboratory result has to include a comment something like
- 17:51Speaker 2: collected through line to elude the ordering provider to the
- 17:54Speaker 2: potential for dilution or contamination that helps them with the
- 17:57Speaker 2: clinical action.
- 17:58Speaker 1: If those three steps aren't for followed, the quality chain
- 18:01Speaker 1: is broken right there. And finally, building on those concepts
- 18:04Speaker 1: of cost utilization and test menus, consider this provocative thought.
- 18:09Speaker 1: When hospital outreach programs prioritize aggressive business growth by facilitating
- 18:13Speaker 1: any test a physician wants, how does that financial incentive
- 18:17Speaker 1: often undermine the clinical goal of limiting unnecessary test utilization
- 18:21Speaker 1: and controlling costs for the rest of the health system.
- 18:24Speaker 2: Something to molover as you integrate these standards of care
- 18:26Speaker 2: into your practice.